Resmed Stellar 150 non-invasive ventilator
MODEL
24162
HSN CODE
90192090
EAN UPC CODE
EAN - NA
Close
Close
Brand:
ResMed
Category:
Device
>
Respiratory Care
DETAILS
Stellar 150 is ResMed's premium non invasive ventilator (with invasive capabilities) for patients with changing ventilatory needs or progressive respiratory conditions, including chronic obstructive pulmonary disease (COPD), neuromuscular disease (NMD) and obesity hypoventilation syndrome (OHS). Easy to set up and use, Stellar 150 carefully strikes a balance between providing clinicians with pre programmed settings that make their job easier, and the opportunity to fine-tune settings when needed to suit each individual patient - a versatile system with its end user in mind. With intelligent backup rate and optional AutoEPAP feature, iVAPS in Stellar 150 continuously monitors ventilation and the upper airway and automatically adjusts to deliver personalised ventilation that accomodates each patient's changing needs
Key Features -
iVAPS
AutoEPAP maintains upper airway patency
intelligent Backup Rate (iBR)
Pathology defaults for obstructive, restrictive, obesity hypoventilation and normal lung mechanics
Invasive and non invasive, adult to peadiatric usage
In built battery back up upto 2 hours
2 customizable pre set programs
Vsync
TiControl
5 Trigger and Cycle Senstivities
Oxygen delivery upto 30 L/min
Comprehensive range of alarms
Real time monitoring and multiple graphs
Integrated SpO2 and FiO2 monitoring
Integrated Humidification
SPECS
Height - 120 mm
Weight - 2.1 kg
Breadth - 170 mm
Length - 230 mm
Sound pressure level - 31 dBA with uncertainty of 3 dBA as measured according to ISO 17510 – 1:2007.
Maximum single fault steady pressure - 60 cm H2O (in all modes)
Breathing resistance under single fault - 2 cm H2O at 30 L/min; 7.2 cm H2O at 60 L/min
Electromagnetic compatibility - Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments.
Aircraft use - Medical-Portable Electronic Devices (M-PED) that meet the Federal Aviation Administration (FAA) requirements of RTCA/DO-160 can be used during all phases of air travel without further testing or approval by the airline operator. ResMed confirms that the Stellar meets RTCA/DO-160 requirements
Storage and transport humidity - 10%–95% non-condensing
Tubing - Flexible plastic, 6’ 6” (2 m) or 9’ 10” (3 m) length (0.9”/22 mm diameter) SlimLine air tubing Flexible plastic, 6’ (1.83 m) length (0.6”/15 mm diameter
Air outlet - 0.9” (22 mm) taper, compatible with ISO 5356-1:2004 Anaesthetic & Respiratory Equipment - Conical Connectors
External DC Power Supply - 24 V, 3 A
Screen Resolution - Dimension (L x W x H): 3” x 2.5” x 0.1” (76.9 mm x 63.9 mm x 3.15 mm) Resolution: 320 x 240 pixels
Operating pressure range - •IPAP: 3 cm H2O to 40 cm H2O (in S, ST, T, PAC mode) •PS: 0 cm H2O to 37 cm H2O (in S, ST, T, PAC mode) •EPAP: 3 cm H2O to 25 cm H2O (in S, ST, T, iVAPS, PAC mode) •CPAP: 4 cm H2O to 20 cm H2O (in CPAP mode only) •Min PS: 0 cm H2O to 20 cm H2O (in iVAPS mode) •Max PS: 0 cm H2O to 30 cm H2O; 8 cm H2O to 30 cm H2O when AutoEPAP is on (in iVAPS mode) •Min EPAP/Max EPAP: 3 cm H2O to 25 cm H2O (in iVAPS mode when AutoEPAP is set to on) Note: iVAPS mode is only available in Stellar 150.
IEC 60601-1 - Class II (Clause 3.14—double insulation). This adherence means the need for an protective earthing (ie, an earthed plug) is not necessary. •Type BF •Continuous operation
Operating humidity - 10%–95% non-condensing
Pressure measurement - Internally mounted flow transducer
Standard air filter - Electro static fibre mesh with TPE frame structure. Bacterial filtration efficiency of 99.540% on area weight 100g/m².
Flow measurement - Internally mounted pressure transducer
Modes - CPAP, S, T, ST with optional iBR, PAC, iVAPS with iBR and optional Auto EPAP
Internal Battery - Lithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh Operating hours: 2 h with a new battery under normal conditions (see below). Patient type: home chronic; pressure: IPAP/EPAP 15/5 cm H2O; mask type: Ultra Mirage; air tubing: 6’ 6” (2 m); leak: 0; respiratory rate: 20 bpm; battery capacity: 100% Patient type: hospital acute; pressure: IPAP/EPAP 20/5 cm H2O; mask type: Ultra Mirage; air tubing: 6’ 6” (2 m); leak: 0; respiratory rate: 45 bpm; battery capacity: 100%
Power supply unit - AC 100–240V, 50–60Hz, 2.2 A, max. 65 W
Storage and transport temperature - -4°F (-20°C) to +140°F (60°C); +122°F (+50°C*)
Housing construction - Flame retardant engineering thermoplastic
Operating temperature - +32°F (0°C) to + 95°F (35°C)
Max FLow - > 200 L/min at 20 cm H2O null
Therapy pressure tolerance - IPAP: ± 0.5 cm H2O ± 10% of set pressure (end of inspiration) EPAP/PEEP: ± 0.5 cm H2O ± 4% of set pressure CPAP: ± 0.5 cm H2O ± 10% of set pressure null
Flow Accuracy - ± 5 L/min or 20% measured value, whichever is greater null
ENQUIRY FORM
Send Enquiry to Manufacturer
Enquire about the price, product information and specification. Enquiry will be sent to Manufacturer and their authorized dealers only.
Full Name
Email Address
Phone Number
Query
Success! Enquiry has been submitted and you will receive further updates over email.
Oops! Something went wrong. Please try again later.
Create Enquiry
Buy Direct
Enquire directly with ResMed and their authorized dealers on the platform
Related products
{{name}}
{{companyName}}