Resmed Stellar 150 non-invasive ventilator

MODEL

24162

HSN CODE

90192090

EAN UPC CODE

EAN - NA

Brand: ResMed
Category: Device > Respiratory Care
 
Stellar 150 is ResMed's premium non invasive ventilator  (with invasive capabilities) for patients with changing ventilatory needs or progressive respiratory conditions, including chronic obstructive pulmonary disease (COPD), neuromuscular disease (NMD) and obesity hypoventilation syndrome (OHS). Easy to set up and use, Stellar 150  carefully strikes a balance between providing clinicians with pre programmed settings that make their job easier, and the opportunity to fine-tune settings when needed to suit each individual patient - a versatile system with its end user in mind. With intelligent backup rate and optional AutoEPAP feature, iVAPS in Stellar 150  continuously monitors ventilation and the upper airway and automatically adjusts to deliver personalised ventilation that accomodates each patient's changing needs 

Key Features -
  • iVAPS
  • AutoEPAP maintains upper airway patency
  • intelligent Backup Rate (iBR)
  • Pathology defaults for obstructive, restrictive, obesity hypoventilation and normal lung mechanics
  • Invasive and non invasive, adult to peadiatric usage
  • In built battery back up upto 2 hours
  • 2 customizable pre set programs
  • Vsync
  • TiControl
  • 5 Trigger and Cycle Senstivities
  • Oxygen delivery upto 30 L/min
  • Comprehensive range of alarms
  • Real time monitoring and multiple graphs
  • Integrated SpO2 and FiO2 monitoring
  • Integrated Humidification
  • Height - 120 mm
  • Weight - 2.1 kg
  • Breadth - 170 mm
  • Length - 230 mm
  • Sound pressure level - 31 dBA with uncertainty of 3 dBA as measured according to ISO 17510 – 1:2007.
  • Maximum single fault steady pressure - 60 cm H2O (in all modes)
  • Breathing resistance under single fault - 2 cm H2O at 30 L/min; 7.2 cm H2O at 60 L/min
  • Electromagnetic compatibility - Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments.
  • Aircraft use - Medical-Portable Electronic Devices (M-PED) that meet the Federal Aviation Administration (FAA) requirements of RTCA/DO-160 can be used during all phases of air travel without further testing or approval by the airline operator. ResMed confirms that the Stellar meets RTCA/DO-160 requirements
  • Storage and transport humidity - 10%–95% non-condensing
  • Tubing - Flexible plastic, 6’ 6” (2 m) or 9’ 10” (3 m) length (0.9”/22 mm diameter) SlimLine air tubing Flexible plastic, 6’ (1.83 m) length (0.6”/15 mm diameter
  • Air outlet - 0.9” (22 mm) taper, compatible with ISO 5356-1:2004 Anaesthetic & Respiratory Equipment - Conical Connectors
  • External DC Power Supply - 24 V, 3 A
  • Screen Resolution - Dimension (L x W x H): 3” x 2.5” x 0.1” (76.9 mm x 63.9 mm x 3.15 mm) Resolution: 320 x 240 pixels
  • Operating pressure range - •IPAP: 3 cm H2O to 40 cm H2O (in S, ST, T, PAC mode) •PS: 0 cm H2O to 37 cm H2O (in S, ST, T, PAC mode) •EPAP: 3 cm H2O to 25 cm H2O (in S, ST, T, iVAPS, PAC mode) •CPAP: 4 cm H2O to 20 cm H2O (in CPAP mode only) •Min PS: 0 cm H2O to 20 cm H2O (in iVAPS mode) •Max PS: 0 cm H2O to 30 cm H2O; 8 cm H2O to 30 cm H2O when AutoEPAP is on (in iVAPS mode) •Min EPAP/Max EPAP: 3 cm H2O to 25 cm H2O (in iVAPS mode when AutoEPAP is set to on) Note: iVAPS mode is only available in Stellar 150.
  • IEC 60601-1 - Class II (Clause 3.14—double insulation). This adherence means the need for an protective earthing (ie, an earthed plug) is not necessary. •Type BF •Continuous operation
  • Operating humidity - 10%–95% non-condensing
  • Pressure measurement - Internally mounted flow transducer
  • Standard air filter - Electro static fibre mesh with TPE frame structure. Bacterial filtration efficiency of 99.540% on area weight 100g/m².
  • Flow measurement - Internally mounted pressure transducer
  • Modes - CPAP, S, T, ST with optional iBR, PAC, iVAPS with iBR and optional Auto EPAP
  • Internal Battery - Lithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh Operating hours: 2 h with a new battery under normal conditions (see below). Patient type: home chronic; pressure: IPAP/EPAP 15/5 cm H2O; mask type: Ultra Mirage; air tubing: 6’ 6” (2 m); leak: 0; respiratory rate: 20 bpm; battery capacity: 100% Patient type: hospital acute; pressure: IPAP/EPAP 20/5 cm H2O; mask type: Ultra Mirage; air tubing: 6’ 6” (2 m); leak: 0; respiratory rate: 45 bpm; battery capacity: 100%
  • Power supply unit - AC 100–240V, 50–60Hz, 2.2 A, max. 65 W
  • Storage and transport temperature - -4°F (-20°C) to +140°F (60°C); +122°F (+50°C*)
  • Housing construction - Flame retardant engineering thermoplastic
  • Operating temperature - +32°F (0°C) to + 95°F (35°C)
  • Max FLow - > 200 L/min at 20 cm H2O null
  • Therapy pressure tolerance - IPAP: ± 0.5 cm H2O ± 10% of set pressure (end of inspiration) EPAP/PEEP: ± 0.5 cm H2O ± 4% of set pressure CPAP: ± 0.5 cm H2O ± 10% of set pressure null
  • Flow Accuracy - ± 5 L/min or 20% measured value, whichever is greater null